Drill Unit & Handpieces (Code: )

Rating:
  • Quantity:

 

 

 

 

DescriptionPart Number
Electric Drill System (Universal) 260-028-150
Motor with Cord 260-028-425
20:1 Mont-Blanc Handpiece
(Single handpiece option when used with Bicon’s electric drill system depicted above.)
260-028-201
18:1 Contra-Angle Handpiece 260-028-181
400:1 Contra-Angle Handpiece 260-028-401

 

DescriptionPart Number
20,000 RPM Air Motor* 260-024-120
Contra-Angle Handpiece Head 260-028-405
Pana Spray** 260-028-500

 

*Not required if using an electric unit.
**Available in USA only (ground shipping).

 

Handpiece Considerations

  • The 400:1 handpiece has sufficient torque to cut through bone at 50 RPM.
  • The 18:1 handpiece is designed to be used with the pilot drill at ~1,100 RPM with external irrigation.
  • Irrigation is not necessary at speeds of 50 RPM or lower.
  • Failure to observe speed limitation may result in burned or overheated bone which could result in substantial bone necrosis.
  • The 20,000 RPM Air Motor attaches to conventional four hole dental tubing and is an alternative to the electric drill unit.
  • Air pressure of at least 80 PSI is recommended to drive the air motor with sufficient torque in dense bone.
  • Proper maintenance of handpieces is crucial for long term success of handpieces. See reference.
  • Do not exceed 132°C (275°F) when sterilizing.
  • In order to prevent discoloration and/or damage to the plating of the handpieces from chemicals that are not sufficiently cleaned from other instruments, do not autoclave the handpieces with other instruments.
  • Always check the handpiece for any abnormal vibration, heating, noise, or sluggish operation. If any abnormality is noticed, cease the use of the handpiece.